Psychrometry: How to Measure Water Potential from the Vapour Above a Leaf
Level: GCSE and A-level Biology, AP Biology and IB Biology, with MDCAT / NEET extension, and first-year university extension marked clearly.
Psychrometry measures the water potential of a sample from the relative humidity of the air in equilibrium with it, in a sealed chamber. Drier samples give drier air. The Kelvin equation turns that humidity into a water potential in MPa.
A leaf disc, a few seeds or a pinch of soil is sealed in a small steel chamber. Water leaves the sample as vapour until the chamber air holds as much vapour as the sample allows. The humidity of that air then reveals the sample's water potential.
This post explains the humidity idea, the Kelvin equation, the thermocouple psychrometer, calibration, and the errors that ruin readings — then tests the whole topic: 20 explained MCQs plus a separate 20-question interactive quiz.
Contents
- Why does psychrometry matter?
- What you will learn
- 1. What is psychrometry?
- 2. Two instruments with one name — the sling psychrometer and the thermocouple psychrometer
- 3. The Kelvin equation — turning humidity into water potential
- 4. The thermocouple psychrometer — the parts
- 5. Two modes — psychrometric and dew-point
- 6. The measurement, step by step
- 7. Calibration — salt solutions of known water potential
- 8. Reading the numbers — ranges and the WP4C
- 9. Measuring solute potential — freeze the tissue first
- 10. What can you measure — leaves, seeds, soil and wood
- 11. Errors and limits
- 12. Psychrometry vs the other methods
- 13. Care and safety — keeping the instrument clean
- 14. Worked examples
- 15. Exam traps — the common mistakes table
- Key points
- 16. Frequently asked questions
- 17. Practice questions — 20 explained MCQs
- 18. Interactive quiz — test yourself
- Continue learning — courses and related posts
- Sources
Why does psychrometry matter?
Water potential decides whether a plant grows, wilts or dies — but you cannot see it. The pressure chamber reads it in a minute, in the field. Psychrometry reads it slowly, in the laboratory — and it can read things the chamber cannot touch.
A pressure chamber needs a whole leaf with a stalk. Psychrometry needs only a disc of tissue, a few seeds, or a pinch of soil. It can measure the water potential of a single small leaf, a seed that must decide whether to germinate, or a soil sample drying in a dish.
It is also the method that reaches the driest samples. A desert seed at −100 MPa is far beyond any pressure chamber — but its vapour pressure still tells the story. Psychrometry reads water potentials from near zero down to −300 MPa.
This post builds the method from zero: the humidity idea, the Kelvin equation, the thermocouple psychrometer, calibration, and the errors that ruin readings.
What you will learn
- What psychrometry is — measuring water potential from the vapour pressure above a sample
- The difference between a sling psychrometer (wet and dry bulb) and a thermocouple psychrometer (sealed chamber)
- The Kelvin equation Ψ = (RT/Vw) ln(RH) — and the humidity values for −1, −1.5 and −8 MPa
- How the Peltier effect cools a thermocouple junction to read dew point
- The two modes — psychrometric (wet-bulb depression) and dew-point (hygrometric)
- How a measurement is made, step by step, and why equilibration takes hours
- Calibration with salt solutions of known water potential
- How to measure solute potential by freezing and thawing tissue
- The method's errors, limits and range — and why temperature control matters most
First-year university extension is marked where it appears. You do not need it for GCSE, A-level, AP, IB, MDCAT or NEET, but it explains why the school rule works.
1. What is psychrometry?
Psychrometry measures the water potential of a sample from the relative humidity of the air in equilibrium with it, in a sealed chamber. Drier samples give drier air. The Kelvin equation turns that humidity into a water potential in MPa.
The idea is simple. Seal a leaf disc in a small chamber and wait. Water leaves the disc as vapour until the air in the chamber holds as much vapour as the disc allows. At that point the air and the disc are in equilibrium — they share the same water potential.
Now measure the humidity of that air. Pure water gives 100% relative humidity. A leaf at −1.5 MPa — the permanent wilting point — gives about 99% humidity. A desert seed at −100 MPa gives about 48%. The drier the sample, the drier its air.
That is the whole method: measure the air, infer the sample. Nothing is squeezed, pressed or cut open during the reading itself. The sample sits undisturbed in its chamber while the instrument reads its vapour.
The method began with the wet-and-dry-bulb psychrometer of meteorology. In 1951, Spanner used the Peltier effect to shrink the idea onto a thermocouple junction small enough to sit inside a sealed sample chamber. That thermocouple psychrometer is the instrument plant physiologists still use.
Exam tip: learn the one-line definition: psychrometry infers water potential from the relative humidity of air in equilibrium with the sample. A question that asks what a psychrometer measures wants "water potential, from vapour pressure".
2. Two instruments with one name — the sling psychrometer and the thermocouple psychrometer
The name "psychrometer" covers two different instruments. The sling psychrometer (wet and dry bulb) measures the humidity of open air. The thermocouple psychrometer measures the water potential of a tissue sample sealed in a chamber. Only the second one belongs in plant physiology.
The sling psychrometer is the older instrument. It has two thermometers: one dry bulb that reads the air temperature, and one wet bulb wrapped in a wet wick. Whirled through the air, water evaporates from the wick and cools the wet bulb.
Dry air evaporates water fast, so the wet bulb cools a lot. Humid air evaporates water slowly, so the wet bulb barely cools. The temperature gap between the two bulbs therefore reads the humidity — a big gap means dry air, a small gap means humid air.
That is a weather instrument. It measures the open air, and it needs moving air to work. It cannot measure a leaf disc, because a leaf disc is not the open air — and its water potential is far too precise a quantity for two glass thermometers.
The thermocouple psychrometer keeps the idea but changes everything else. The sample is sealed in a steel chamber with a thermocouple junction. There is no whirling and no wick. Instead, an electric current cools the junction below the dew point. Water condenses on it, and the evaporation cools the junction by an amount that depends on the chamber's humidity.
The two instruments share a word and a principle — evaporation cooling reads humidity. They do not share a job. If an exam asks which psychrometer measures leaf water potential, the answer is always the thermocouple psychrometer.
Exam tip: learn the pair: sling psychrometer = air humidity; thermocouple psychrometer = tissue water potential. A question that asks why a sling psychrometer cannot measure a leaf wants "it measures open air, not a sealed sample".
3. The Kelvin equation — turning humidity into water potential
The Kelvin equation links water potential to relative humidity: Ψ = (RT/Vw) ln(RH). At 25 °C, Ψ = 137.7 × ln(RH) MPa. A leaf at −1.5 MPa sits at about 99% humidity; a desert seed at −100 MPa sits at about 48%.
The equation has four parts. R is the gas constant (8.314 J mol⁻¹ K⁻¹). T is the absolute temperature in kelvin. Vw is the molar volume of water (18 × 10⁻⁶ m³ mol⁻¹). RH is the relative humidity as a fraction — 0.99, not 99%.
At 25 °C (298 K), the factor RT/Vw works out to about 137.7 MPa. So the working form is simple: water potential in MPa is 137.7 times the natural logarithm of the relative humidity. The logarithm is negative whenever RH is below 1 — which is why water potentials come out negative.
The table below gives the humidity for the water potentials that matter most. Read it in both directions: a humidity reading gives a water potential, and a water potential predicts a humidity.
| Water potential (MPa) | Relative humidity at 25 °C | What it is |
|---|---|---|
| 0 | 100% | Pure water |
| −0.03 | 99.98% | Field capacity (wet soil) |
| −0.1 | 99.93% | A well-watered leaf |
| −0.5 | 99.64% | A mildly stressed leaf |
| −1.0 | 99.28% | A stressed crop leaf |
| −1.5 | 98.92% | Permanent wilting point |
| −8 | 94.4% | Lower limit of water uptake |
| −10 | 93.0% | A very dry leaf |
| −100 | 48% | A dry seed |
| −300 | 11% | An extremely dry sample |
The sample behind these numbers is usually a leaf disc, a few seeds or a pinch of soil, sealed in a small steel chamber. The disc gives up vapour until the chamber air is in balance with it — and that balance is what the table describes.
Look at the top of the table. Between 0 and −1.5 MPa — the whole range of a living, growing plant — the humidity only moves from 100% to 99%. That is why psychrometry is so hard: the instrument must tell 99.93% from 99.28% apart, a difference of half a percent.
Look at the bottom instead. Below −10 MPa the humidity falls fast — 93% at −10 MPa, 48% at −100 MPa, 11% at −300 MPa. Dry samples are easy to read. Wet samples are the hard ones, and that is the method's central weakness.
First-year university extension: the Kelvin equation comes from the chemical potential of water. Water moves from high to low chemical potential, and the vapour above a sample carries the same potential as the sample itself. The logarithm appears because chemical potential depends on the logarithm of vapour pressure — the same reason pH is logarithmic.
Exam tip: learn two anchor pairs: −1.5 MPa ≈ 99% RH (wilting point) and −8 MPa ≈ 94% RH (limit of uptake). A question that asks for the humidity above a leaf at −1.5 MPa wants "about 99%".
4. The thermocouple psychrometer — the parts
A thermocouple psychrometer has four working parts: a sealed sample chamber, a sensing thermocouple junction, a dry reference junction, and a microvoltmeter or datalogger. A current cools the sensing junction below the dew point, and the voltage that follows reads the humidity.
- The sample chamber. A small stainless-steel chamber — the classic Wescor C-52 — that seals gas-tight around the sample. A leaf disc, a few seeds or a pinch of soil goes inside. The chamber's job is to hold the sample and its air together until they equilibrate.
- The sensing junction. A tiny thermocouple junction inside the chamber, just above the sample. It is both the cooler and the thermometer: a current through it cools it by the Peltier effect, and the voltage across it reports its temperature.
- The reference junction. A second, dry junction that stays at the chamber's temperature. All readings are the difference between the sensing junction and this reference — so any drift in room temperature cancels out.
- The microvoltmeter or datalogger. The classic readout is the Wescor HR-33T dew point microvoltmeter; modern work uses a datalogger, which supplies the cooling current and records the microvolt output.
The thermocouple is the heart of the instrument, and it is small for a reason. A junction the size of a pinhead cools and warms in seconds, and it disturbs almost no air. A big sensor would carry heat into the chamber and wreck the reading.
The chamber is small for the same reason. A small air space equilibrates with the sample faster than a large one, and it holds less water vapour to confuse the reading. Most sample chambers hold only a few millilitres of air.
The microvoltmeter reading means nothing on its own. It becomes a water potential only through the psychrometer's calibration — a straight-line plot of microvolt output against known water potentials, made from salt solutions. Section 7 explains how that line is made.
First-year university extension: a thermocouple is two different metals joined together. A temperature difference across the join makes a voltage (the Seebeck effect), and a current through the join makes a temperature difference (the Peltier effect). The psychrometer uses both: Peltier cooling to make dew, Seebeck voltage to read the result.
Exam tip: learn the parts list: sample chamber, sensing junction, reference junction, microvoltmeter. A question that asks what the Peltier effect does in a psychrometer wants "it cools the sensing junction below the dew point".
5. Two modes — psychrometric and dew-point
A thermocouple psychrometer works in two modes. In psychrometric mode it measures the wet-bulb depression after Peltier cooling. In dew-point (hygrometric) mode it measures the dew-point depression. Dew-point mode is about twice as sensitive and less troubled by temperature.
Both modes start the same way. A brief current — typically about 8 milliamperes — cools the sensing junction below the dew point of the chamber air. Water condenses on the cold junction as a tiny drop of dew.
In psychrometric mode, the current is switched off and the dew evaporates. Evaporation cools the junction below the chamber temperature, and the size of that cooling — the wet-bulb depression — depends on the humidity. Dry chamber air evaporates the dew fast and cools the junction a lot; humid air cools it only a little.
In dew-point mode, the instrument holds the junction exactly at the dew point — the temperature where dew just forms and just evaporates. The gap between the chamber temperature and the dew point — the dew-point depression — reads the humidity directly. Drier air has a lower dew point, so the gap is bigger.
Dew-point mode wins on two counts. Its voltage sensitivity is about 7 microvolts per MPa at 25 °C, against about 4 microvolts per MPa for psychrometric mode — nearly twice the signal for the same water potential. It is also less sensitive to temperature drift, which is the method's worst enemy.
Most modern instruments, including the HR-33T, can run in either mode. Dew-point mode is the usual choice for plant tissue. Psychrometric mode survives mainly for soil work and for older datalogger programs.
Exam tip: learn the pair: psychrometric mode = wet-bulb depression; dew-point mode = dew-point depression. A question that asks which mode is more sensitive wants "dew-point, about 7 µV per MPa".
6. The measurement, step by step
The measurement has six steps: cut the sample, seal it in the chamber, wait for vapour equilibrium, cool the junction, read the microvolts, and convert to MPa with the calibration. The waiting is the longest part — equilibration often takes one to four hours.
- Cut the sample. Punch a leaf disc with a cork borer, or take a few seeds or a pinch of soil. Handle the sample with clean forceps, never fingers — skin oils contaminate the chamber and ruin the reading.
- Seal it in the chamber. Place the sample in the steel chamber and close it gas-tight. The sensing junction sits just above the sample, not touching it. Any contact with liquid water gives a false reading of zero.
- Wait for equilibrium. Put the sealed chamber in a water bath or an insulated box at a steady temperature — usually 25 °C. Water vapour leaves the sample until the chamber air is in balance with it. This takes one to four hours for most samples.
- Cool the junction. The datalogger passes a brief current (about 8 mA) through the sensing junction. Peltier cooling drops it below the dew point, and a drop of dew condenses on it.
- Read the output. In dew-point mode the instrument holds the junction at the dew point and records the voltage. In psychrometric mode it records the wet-bulb depression as the dew evaporates. The output is a few microvolts.
- Convert to water potential. The microvolt reading is converted to MPa using the psychrometer's own calibration factor (Section 7). Repeat the reading every 30 to 60 minutes until two or three in a row agree — that is when equilibrium is confirmed.
Equilibration is the step students underestimate. A leaf disc is mostly water, so it gives up vapour quickly — but the chamber walls, the junction and the air itself must all reach the same temperature first. Any temperature difference drives vapour one way or the other, and the reading drifts.
That is why the water bath matters. The whole chamber sits in water at a fixed temperature, so the sample and the sensor stay equal to within a thousandth of a degree. A chamber left on a bench in a draughty room never settles.
Exam tip: learn the sequence: cut, seal, equilibrate, cool, read, convert. A question that asks why a psychrometer reading takes hours wants "the sample and the chamber air must reach vapour equilibrium at a steady temperature".
Watch a measurement, stage by stage. One leaf disc, from sealing to the final reading.
7. Calibration — salt solutions of known water potential
Every thermocouple psychrometer is calibrated against salt solutions of known water potential — usually 4 to 6 solutions of KCl or NaCl. A disc of filter paper soaked in each solution is measured in turn, and the microvolt readings are plotted against the known values to give the calibration factor.
Calibration is needed because no two thermocouples are identical. Tiny differences in how each junction was made change its voltage output, and the output drifts with use. A factory calibration would be wrong within weeks — so each psychrometer carries its own calibration, made by its user.
The standards are simple salt solutions. Dissolving salt lowers the water potential of water by a known amount — the solute potential, given by van't Hoff's equation. A 0.1 molal NaCl solution has a water potential of about −0.46 MPa at 25 °C; stronger solutions are more negative, as the table shows.
| NaCl solution (molal) | Water potential at 25 °C (MPa) |
|---|---|
| 0.1 | About −0.46 |
| 0.3 | About −1.38 |
| 0.5 | About −2.29 |
| 1.0 | About −4.62 |
The procedure is the same as a sample measurement. A disc of filter paper is soaked in one salt solution and sealed in the chamber. After equilibration — 30 minutes to 6 hours depending on the solution and the chamber — the microvolt output is recorded. The process repeats for each solution.
The known water potentials are then plotted against the microvolt readings. The result is a straight line, and its slope is the calibration factor for that psychrometer — the number that turns future microvolt readings into MPa. The line is made at one temperature, usually 25 °C, and the psychrometer must be used at that same temperature.
First-year university extension: 25 °C is the standard calibration temperature because it gives the lowest variability — Brown and Bartos showed this in 1982, comparing 15, 25 and 35 °C. It is also practical: a water bath at 25 °C in a warm laboratory needs only a heater, not a cooler.
Exam tip: learn the calibration in one line: "known salt solutions, filter paper disc, plot µV against known Ψ". A question that asks why each psychrometer needs its own calibration wants "junctions differ, so outputs differ".
8. Reading the numbers — ranges and the WP4C
A Peltier thermocouple psychrometer reads reliably from about −0.3 to −8 MPa. Near zero it cannot resolve the humidity, and below −8 MPa its signal is too small. The chilled-mirror WP4C reaches −300 MPa and is the laboratory standard for dry samples.
The wet end is the problem. At −0.1 MPa the chamber air is at 99.93% humidity; at −0.03 MPa it is at 99.98%. No thermocouple can tell those apart — the voltages differ by less than the electrical noise. So the upper limit is about −0.03 to −0.2 MPa. Wetter samples need a tensiometer or a pressure chamber instead.
The dry end is kinder. Down to about −8 MPa the signal stays clean, and wet-loop sensors reach −300 MPa. The reliable working range quoted for plant and soil work is −0.3 to −8 MPa — wide enough for wilting leaves, dry soil and most seeds.
The modern laboratory instrument is the WP4C dew point potentiometer. It replaces the thermocouple with a chilled mirror: a tiny mirror is cooled until dew just forms on it, and the mirror and sample temperatures are measured to 0.001 °C.
Its range is 0 to −300 MPa, with an accuracy of ±0.05 MPa from 0 to −5 MPa and 1% from −5 to −300 MPa.
The WP4C is fast by psychrometer standards — about 10 to 15 minutes for a soil sample in precise mode, about 20 minutes for plant tissue, under 5 minutes in fast mode.
It is so accurate that it is used to calibrate other methods. But it shares the wet-end limit: it cannot measure accurately near field capacity (−0.033 MPa), where a reading of −0.1 MPa carries an error of ±50%.
Temperature sets the accuracy of every range above. A gap of only 0.006 °C between sample and sensor is enough to use up the WP4C's ±0.05 MPa accuracy — which is why the instrument measures both temperatures to 0.001 °C.
Exam tip: learn the range pair: thermocouple psychrometer −0.3 to −8 MPa; WP4C 0 to −300 MPa. A question that asks which instrument reads a dry seed at −100 MPa wants "the WP4C chilled-mirror potentiometer".
9. Measuring solute potential — freeze the tissue first
Freezing and thawing a leaf disc destroys its membranes, so turgor falls to zero. A psychrometer reading on the thawed disc is then the solute potential alone. Subtracting it from the total water potential gives the turgor pressure.
A normal psychrometer reading gives the total water potential: Ψw = Ψs + Ψp. That is one number hiding two. To separate them, the tissue is frozen — in a freezer or in liquid nitrogen — and then thawed.
Freezing bursts the cell membranes. The solutes that were locked inside the vacuole spill into the cell wall space, and the pressure that the wall once held — the turgor — is gone. Thawed tissue has Ψp = 0, so whatever the psychrometer reads is Ψs alone.
The arithmetic is then simple. Measure the fresh disc (Ψw), measure the frozen-thawed disc (Ψs), and subtract: Ψp = Ψw − Ψs. A leaf at Ψw = −1.2 MPa with Ψs = −1.8 MPa has a turgor of +0.6 MPa — the number the pressure probe would read directly.
The thawed disc is measured exactly like a fresh one: sealed in the chamber, equilibrated, and read the same way. Only the tissue has changed — its membranes are gone, so its reading is solute potential, not total water potential.
There is a second way to get solute potential: isopiestic psychrometry. The sample is sealed in a chamber with a salt solution of known water potential, and the sample is weighed before and after.
If it gains weight it was wetter than the solution; if it loses weight it was drier. The solution that leaves its weight unchanged has the sample's water potential.
Exam tip: learn the freeze-thaw rule: frozen-thawed tissue gives Ψs, because Ψp = 0. A question that asks how to find turgor from psychrometer readings wants "Ψp = Ψw − Ψs".
10. What can you measure — leaves, seeds, soil and wood
Psychrometry measures anything small enough to seal in a chamber: leaf discs, seeds, soil, wood blocks and stem segments. It also works in place — leaf and stem psychrometers clip onto living plants, and soil sensors sit buried for months.
Leaves are the commonest sample. A disc punched from a leaf gives the leaf's water potential without a stalk, which is why psychrometry suits small leaves — grasses, seedlings and herbs — that a pressure chamber cannot hold. Several discs from one leaf can be averaged, just as several leaves are averaged for a chamber reading.
Seeds are the method's showcase. A dry seed at −50 or −100 MPa is unreadable by any pressure method, but its humidity is easy to read — 48% at −100 MPa. Seed banks use psychrometry to check seeds are dry enough to keep, and germination studies use it to find the water potential at which a seed sprouts.
Soil is the second home of psychrometry. A pinch of soil in a chamber gives its total water potential — matric plus osmotic together. Caged-screen sensors buried in the ground do the same job for months, logging soil drying through a whole season without anyone digging the sample up.
Wood and stem segments work too — blocks too big or too woody for a pressure chamber fit a psychrometer chamber easily. And in-situ psychrometers clip onto a living stem or leaf, sealing a small patch of tissue against the sensor so the plant is measured while it is still growing.
Drought screening is where the method earns its keep. Breeders grow dozens of wheat or maize varieties in dry soil and measure one leaf disc from each. The variety that holds its water potential highest — the least negative — is the one that keeps growing when the rain stops.
Exam tip: learn the sample list: leaf discs, seeds, soil, wood, stem segments. A question that asks what psychrometry can measure that a pressure chamber cannot wants "seeds and soil at very negative water potentials".
11. Errors and limits
Psychrometry's worst enemy is temperature. A 1 °C difference between the sample and the sensor can cause an error of more than 10 MPa. Everything else — contamination, slow equilibration, damaged cells — matters less, but all of it matters.
- Temperature differences. The Kelvin equation is steep: a tiny temperature gap between sample and sensor moves vapour one way and shifts the reading by megapascals. Early workers controlled temperature to ±0.001 °C to reach 0.01 MPa accuracy. The WP4C needs temperature differences accurate to 0.006 °C for its ±0.05 MPa accuracy.
- Slow equilibration. A reading taken before the chamber air is in balance with the sample is a reading of nothing. Leaf discs need one to four hours; soil can need longer. Impatience is the commonest student error.
- Contamination. Skin oils, sap residues and dried salt on chamber walls all give off or absorb vapour. Every surface inside the chamber must be cleaned between samples, and the junctions cleaned and recalibrated regularly.
- Damaged cells. Cutting a leaf disc wounds cells, and wounded cells leak solutes. The leaked solutes raise the sample's water potential inside the chamber, and respiration during equilibration adds water too. Both push the reading toward zero.
- Water on the walls. Vapour can adsorb onto the chamber walls, especially in humid chambers. The walls then act as a second sample, and the reading is a mixture of the two.
- The wet-end limit. Near zero the humidity differences are too small to resolve (Section 8). A well-watered leaf at −0.05 MPa is barely measurable — the pressure chamber does that job better.
- One reading, hours of waiting. A pressure chamber gives a reading in a minute. A psychrometer gives one in hours. Field surveys of a hundred plants belong to the chamber; the psychrometer belongs to the laboratory.
The temperature problem deserves one more line. The chamber air at 99% humidity holds very little water — so little that warming the sample by a fraction of a degree drives enough vapour off it to change the reading by a megapascal. That is why the water bath is not optional equipment. It is the instrument.
First-year university extension: the same temperature sensitivity explains why psychrometer and pressure chamber readings sometimes disagree. The psychrometer detects solutes in the cell walls (apoplastic solutes) that the chamber misses, and the chamber can read lower when damaged cells leak solutes into the psychrometer chamber. Boyer compared the two methods in 1967; later work confirmed the apoplastic difference.
Exam tip: learn the error list as "temperature, time, contamination, damaged cells". A question that asks for the biggest source of error wants "temperature differences between sample and sensor".
12. Psychrometry vs the other methods
Five methods measure plant water relations, and each measures something different. The pressure chamber reads leaf water potential in the field in a minute. Psychrometry reads it in the laboratory in hours — but from samples the chamber cannot touch, and down to −300 MPa.
| Method | What it measures | Scale | Strengths | Limits |
|---|---|---|---|---|
| Pressure chamber | Leaf / shoot water potential | Whole leaf | Fast (1 min), portable, field-ready | Needs a stalked leaf; stops near −8 MPa; destructive |
| Thermocouple psychrometer | Sample water potential (from vapour) | Leaf disc, seed, soil | Small samples; seeds and soil; −0.3 to −8 MPa | Slow (hours); very temperature-sensitive; lab method |
| WP4C chilled mirror | Sample water potential (from dew point) | Leaf, seed, soil cup | Fast (5–20 min); 0 to −300 MPa; ±0.05 MPa | Poor near field capacity; laboratory only |
| Cell pressure probe | Turgor pressure directly | One living cell | Direct; live readings as conditions change | Delicate; microscope work; laboratory only |
| Incipient plasmolysis | Cell solute potential (Ψp = 0) | Tissue sample | Simple school practical; cheap | Slow; cells adjust solutes during the assay |
Choose the method by the question. For "does this vineyard need water today?" the pressure chamber wins — it is fast and it works in the field. For "how dry is this wheat seed?" only a vapour method will do. For "what is the turgor of this one guard cell?" only the pressure probe will do.
Psychrometry and the pressure chamber usually agree — but not always. The psychrometer reads the whole tissue including solutes in the cell walls, which the chamber cannot see.
And a disc cut for the psychrometer leaks solutes from wounded cells, which can push its reading toward zero. When the two disagree, both readings are honest — they are measuring slightly different things.
Exam tip: learn the method map: chamber = leaf Ψ in the field; psychrometer = small-sample Ψ in the lab; WP4C = dry-sample Ψ; probe = single-cell turgor; plasmolysis = cell Ψs. A question that asks which method reads a seed at −100 MPa wants "chilled-mirror psychrometry (WP4C)".
13. Care and safety — keeping the instrument clean
Psychrometry has no high pressure and no dangerous gas, so its safety rules are about care, not injury. Clean chambers, clean junctions, clean hands — contamination is the instrument's worst enemy after temperature.
- Clean between samples. Wipe every chamber surface to remove sap, salt residue and dust. A dirty chamber remembers the last sample and mixes it into the next reading.
- Handle samples with forceps. Fingers leave skin oils on leaf discs and chamber walls. Oils change how water condenses on the junction and shift the reading.
- Clean the junctions. The thermocouple junctions collect deposits with use. Clean them exactly as the manufacturer directs, then recalibrate — cleaning changes the calibration.
- Use distilled water only. The wet wick of a sling psychrometer — and any water used near the instrument — must be distilled. Tap water leaves salts behind when it evaporates.
- Keep the water bath steady. Check the bath temperature before a run of readings. A bath that drifted overnight invalidates every reading taken in it.
- Store chambers dry. A damp chamber grows mould, and mould gives off vapour of its own. Dry and seal chambers between uses.
Compared with the pressure chamber, psychrometry is gentle work — no cylinder of compressed nitrogen, no steel vessel under pressure. The risk is not to the worker but to the data: a careless reading looks exactly like a careful one, and only the calibration reveals the difference.
Exam tip: learn the care list as "clean, forceps, distilled water, steady bath". A question that asks why forceps are used wants "fingers leave skin oils that contaminate the chamber".
14. Worked examples
All numbers in this section are practice figures, invented for teaching. They are not measurements from a real experiment. Use them to practise the method, not to cite as data.
Worked example 1 — From humidity to water potential. A leaf disc equilibrates in a psychrometer chamber at 25 °C. The chamber air is at 99.28% relative humidity. What is the leaf's water potential?
Use the Kelvin equation at 25 °C: Ψ = 137.7 × ln(RH). RH as a fraction is 0.9928. ln(0.9928) = −0.00723. So Ψ = 137.7 × (−0.00723) ≈ −1.0 MPa.
The leaf is stressed — about the same as a grapevine at the top of its mild-stress band. A grower would watch it, but not irrigate yet.
Worked example 2 — From microvolts to water potential. A dew-point psychrometer has a sensitivity of 7.5 µV per MPa at 25 °C. A leaf disc gives a reading of −11.25 µV. What is its water potential? The same disc, frozen and thawed, reads −1.8 MPa. What is the turgor pressure?
Water potential is the reading divided by the sensitivity: −11.25 ÷ 7.5 = −1.5 MPa. That is the total water potential, Ψw — the permanent wilting point.
The frozen-thawed reading is the solute potential: Ψs = −1.8 MPa. Turgor is Ψp = Ψw − Ψs = −1.5 − (−1.8) = +0.3 MPa. The leaf is at wilting point with only a little turgor left — it will wilt by midday.
Worked example 3 — Calibrating with a salt solution. A new thermocouple psychrometer is calibrated with a 0.3 molal NaCl solution (Ψ = −1.38 MPa at 25 °C). The reading is −10.35 µV. What is the calibration factor in µV per MPa? A leaf disc then reads −6.9 µV. What is its water potential?
The calibration factor is the reading divided by the known potential: −10.35 ÷ −1.38 = 7.5 µV per MPa. That is the dew-point sensitivity at 25 °C.
The leaf reading converts with the same factor: −6.9 ÷ 7.5 = −0.92 MPa. The leaf is mildly stressed — inside the range where a pressure chamber would agree closely.
15. Exam traps — the common mistakes table
| Trap | What students write | What is correct |
|---|---|---|
| Which psychrometer | "A sling psychrometer measures leaf water potential" | Only the thermocouple psychrometer does. The sling measures air humidity. |
| What is measured | "Psychrometry measures water content" | It measures water potential — from vapour pressure, not water content. |
| The equation | "Ψ = RH × 100" | Ψ = (RT/Vw) ln(RH). At 25 °C, Ψ = 137.7 × ln(RH) MPa. |
| Humidity values | "A leaf at −1.5 MPa sits at 50% humidity" | It sits at about 99% humidity. 50% humidity means about −95 MPa. |
| The modes | "Psychrometric and dew-point modes are the same" | Psychrometric reads wet-bulb depression; dew-point reads dew-point depression. |
| Sensitivity | "Both modes are equally sensitive" | Dew-point is about twice as sensitive — about 7 µV per MPa vs about 4. |
| Non-destructive | "Psychrometry is fully non-destructive" | The disc is cut from the plant. The reading itself does not damage the disc further. |
| Temperature | "A degree or two of temperature drift does not matter" | A 1 °C sample-sensor gap can cause an error over 10 MPa. |
| Equilibration | "Read the psychrometer as soon as the sample is sealed in" | Wait one to four hours until two or three readings in a row agree. |
| Calibration | "One calibration works for every psychrometer" | Every psychrometer needs its own calibration — junctions differ. |
| Wet end | "Psychrometry is best for well-watered leaves" | Near zero it cannot resolve the humidity. The chamber does wet leaves better. |
| Freeze-thaw | "A frozen-thawed disc gives total water potential" | It gives solute potential — freezing destroys membranes, so turgor is zero. |
| The Peltier effect | "The Peltier effect heats the junction" | A current cools the junction below the dew point so water condenses on it. |
| Who invented it | "Scholander invented the psychrometer in 1965" | Spanner built the thermocouple psychrometer in 1951. Scholander built the pressure chamber. |
Key points
- Psychrometry: water potential inferred from the relative humidity of air in equilibrium with the sample, in a sealed chamber.
- Two instruments: sling psychrometer (wet and dry bulb) measures air humidity; thermocouple psychrometer measures tissue water potential. Only the second belongs in plant physiology.
- Kelvin equation: Ψ = (RT/Vw) ln(RH). At 25 °C, Ψ = 137.7 × ln(RH) MPa.
- Anchor values: −1.5 MPa ≈ 99% RH (wilting point); −8 MPa ≈ 94% RH (limit of uptake); −100 MPa ≈ 48% RH (dry seed).
- History: Spanner built the thermocouple psychrometer in 1951 using the Peltier effect; Richards and Ogata refined it in 1958; the Wescor HR-33T and C-52 chamber became the standard.
- Parts: sealed sample chamber, sensing junction, dry reference junction, microvoltmeter or datalogger.
- Two modes: psychrometric (wet-bulb depression, about 4 µV per MPa) and dew-point (about 7 µV per MPa, less temperature-sensitive).
- Procedure: cut the sample, seal it in, equilibrate 1–4 hours at steady temperature, cool the junction, read microvolts, convert with the calibration.
- Calibration: 4–6 KCl or NaCl solutions of known Ψ (0.1 molal NaCl ≈ −0.46 MPa at 25 °C); each psychrometer needs its own calibration, at the temperature it will be used.
- Range: Peltier psychrometer about −0.3 to −8 MPa; WP4C chilled mirror 0 to −300 MPa, ±0.05 MPa from 0 to −5 MPa.
- Solute potential: freeze and thaw the tissue (Ψp = 0), then read Ψs; turgor is Ψp = Ψw − Ψs.
- Samples: leaf discs, seeds, soil, wood, stem segments; in-situ leaf and stem psychrometers measure living plants.
- Biggest error: temperature — a 1 °C sample-sensor gap can cause an error over 10 MPa. Next: slow equilibration, contamination, damaged cells.
- The one-line summary: seal the sample in, wait for its vapour to balance, read the humidity — and the Kelvin equation turns that humidity into a water potential.
16. Frequently asked questions
What is psychrometry in simple words?
It measures how dry a leaf is from the humidity of the air above it. A leaf disc is sealed in a chamber, and the humidity of the air above it is measured. Drier leaves make drier air — and the Kelvin equation turns that humidity into a water potential.
What is the difference between a sling psychrometer and a thermocouple psychrometer?
The sling psychrometer has two thermometers — wet bulb and dry bulb — and measures the humidity of open air. The thermocouple psychrometer seals a tissue sample in a chamber and measures the humidity of the air in equilibrium with that sample. Only the thermocouple version measures water potential.
What is the Kelvin equation?
It is the link between water potential and humidity: Ψ = (RT/Vw) ln(RH). At 25 °C it becomes Ψ = 137.7 × ln(RH) MPa. Put in the relative humidity as a fraction, and the water potential in MPa comes out.
Why does a psychrometer reading take hours?
The sample and the air in the chamber must reach vapour equilibrium at a perfectly steady temperature. That takes one to four hours for most samples. A reading taken early is a reading of air that is still changing.
Why is temperature control so important?
Because the humidity differences being measured are tiny — 99.93% versus 99.28%. A temperature gap of 1 °C between the sample and the sensor can cause an error of more than 10 MPa. Chambers sit in a water bath at a fixed temperature, usually 25 °C.
Can psychrometry measure a well-watered leaf?
Badly. Near zero the humidity is 99.9% or more, and no psychrometer can resolve the differences there. The practical limit is about −0.03 to −0.2 MPa. Well-watered leaves are measured with a pressure chamber or a tensiometer instead.
How does freezing a leaf help measure solute potential?
Freezing bursts the cell membranes, so turgor falls to zero. The psychrometer then reads the solute potential alone, because Ψw = Ψs when Ψp = 0. Subtracting that from the fresh reading gives the turgor pressure.
How is psychrometry different from the pressure chamber?
The chamber squeezes a whole leaf in the field and reads in a minute; psychrometry listens to vapour in the laboratory and reads in hours. Psychrometry works on leaf discs, seeds and soil the chamber cannot touch, and reaches −300 MPa — but it cannot match the chamber for speed or for wet leaves.
17. Practice questions — 20 explained MCQs
These are original practice questions, written for this post. They are not past-paper reproductions, and they are not claimed to be past-paper questions.
Work through them without looking at the answers first, then check each explanation. If you get one wrong, re-read the section it tests.
Tap each question to reveal its answer and explanation.
MCQ 1. What does psychrometry measure?
A. Water potential, from vapour pressure.
B. Water content of the leaf.
C. Turgor pressure directly.
D. Photosynthesis rate.
Answer: A
Explanation: The humidity of air in equilibrium with the sample gives Ψ. Content is not potential, and turgor needs a pressure probe.
MCQ 2. Which instrument measures the water potential of a leaf disc?
A. Sling psychrometer.
B. Thermocouple psychrometer.
C. Hair hygrometer.
D. Tensiometer.
Answer: B
Explanation: It seals the sample in a chamber with a thermocouple. The sling measures open-air humidity, and a tensiometer measures soil matric potential only.
MCQ 3. The Kelvin equation is:
A. Ψ = RH × 100.
B. Ψ = P + Ψs.
C. Ψ = (RT/Vw) ln(RH).
D. Ψ = −P.
Answer: C
Explanation: At 25 °C, Ψ = 137.7 × ln(RH) MPa. Ψ = −P is the pressure-chamber conversion.
MCQ 4. A leaf at the permanent wilting point (−1.5 MPa) sits at about what humidity?
A. 50%.
B. 90%.
C. 100%.
D. 99%.
Answer: D
Explanation: −1.5 MPa is 98.92% RH at 25 °C. 50% humidity means about −95 MPa.
MCQ 5. In a thermocouple psychrometer, the sensing junction is cooled by:
A. The Peltier effect.
B. Blowing air over it.
C. Liquid nitrogen.
D. An ice bath.
Answer: A
Explanation: A current through the junction cools it below the dew point. The chamber is sealed, so no air is blown over it.
MCQ 6. Dew-point mode is preferred over psychrometric mode because it is:
A. Cheaper to build.
B. About twice as sensitive and less temperature-sensitive.
C. Faster to equilibrate.
D. Able to measure single cells.
Answer: B
Explanation: About 7 µV per MPa vs about 4. Equilibration time is the same in both modes.
MCQ 7. Why must the sample and the chamber air reach equilibrium before reading?
A. To warm the sample.
B. To dry the sample.
C. So the air humidity matches the sample's water potential.
D. To kill the leaf.
Answer: C
Explanation: The reading means nothing before equilibrium. Drying would change the sample.
MCQ 8. A thermocouple psychrometer is calibrated with:
A. Distilled water only.
B. Other psychrometers.
C. Pressure-chamber readings.
D. Salt solutions of known water potential.
Answer: D
Explanation: Usually 4–6 KCl or NaCl solutions. Pure water gives only one point, Ψ = 0.
MCQ 9. A 0.1 molal NaCl solution at 25 °C has a water potential of about:
A. −0.46 MPa.
B. +0.46 MPa.
C. −4.6 MPa.
D. 0 MPa.
Answer: A
Explanation: Dissolved salt lowers the water potential. −4.6 MPa is a 1.0 molal solution.
MCQ 10. What is the reliable working range of a Peltier thermocouple psychrometer?
A. 0 to −0.03 MPa.
B. About −0.3 to −8 MPa.
C. 0 to −300 MPa.
D. −8 to −300 MPa.
Answer: B
Explanation: Wetter samples cannot be resolved; drier ones give too little signal. 0 to −300 MPa is the WP4C.
MCQ 11. The WP4C measures water potential with:
A. A pressure transducer.
B. A sling psychrometer.
C. A chilled mirror.
D. A glass electrode.
Answer: C
Explanation: The mirror is cooled until dew forms on it. It uses vapour pressure, not a pressure transducer.
MCQ 12. The WP4C is most accurate in which range?
A. Near field capacity.
B. 0 to −0.03 MPa.
C. −100 to −300 MPa.
D. 0 to −5 MPa, at ±0.05 MPa.
Answer: D
Explanation: From −5 to −300 MPa the accuracy is 1%. Near field capacity it cannot measure accurately.
MCQ 13. Freezing and thawing a leaf disc before psychrometry gives:
A. Solute potential.
B. Total water potential.
C. Turgor pressure.
D. Matric potential.
Answer: A
Explanation: Membranes break, turgor falls to zero, so Ψ = Ψs. Turgor is found by subtraction.
MCQ 14. A fresh leaf disc reads Ψw = −1.2 MPa. The frozen-thawed disc reads Ψs = −1.8 MPa. The turgor is:
A. −3.0 MPa.
B. +0.6 MPa.
C. −0.6 MPa.
D. 0 MPa.
Answer: B
Explanation: Ψp = Ψw − Ψs = −1.2 − (−1.8) = +0.6 MPa. Adding the two values gives −3.0 MPa.
MCQ 15. Which sample is hardest for a vapour-pressure psychrometer to measure?
A. A dry seed at −100 MPa.
B. Dry soil at −50 MPa.
C. A well-watered leaf near −0.05 MPa.
D. A wilted leaf at −3 MPa.
Answer: C
Explanation: Near zero the humidity is 99.97% or more. Dry samples are easy to read.
MCQ 16. A 1 °C temperature difference between sample and sensor can cause an error of:
A. 0.01 MPa.
B. 0.1 MPa.
C. 1 MPa.
D. More than 10 MPa.
Answer: D
Explanation: Temperature is the method's worst enemy. 0.01 MPa needs control to ±0.001 °C.
MCQ 17. Who built the first thermocouple psychrometer, in 1951?
A. Spanner.
B. Scholander.
C. Dixon.
D. Höfler.
Answer: A
Explanation: He used the Peltier effect to measure suction pressure. Scholander built the pressure chamber in 1965; Dixon built an early glass pressure chamber in 1922.
MCQ 18. Why are samples handled with forceps, not fingers?
A. Fingers are too warm.
B. Skin oils contaminate the chamber.
C. Fingers crush the sample.
D. Forceps are sterile.
Answer: B
Explanation: Oils change condensation on the junction. Cleanliness, not sterility, is the point.
MCQ 19. Psychrometry and the pressure chamber sometimes disagree because:
A. They use different units.
B. The chamber is always wrong.
C. The psychrometer detects cell-wall solutes the chamber misses.
D. Psychrometers only work at night.
Answer: C
Explanation: Apoplastic solutes affect the vapour reading only. Both methods report MPa, and both readings are honest.
MCQ 20. The main advantage of psychrometry over the pressure chamber is that it can measure:
A. Whole leaves faster.
B. Single cells directly.
C. Wet leaves more accurately.
D. Small samples such as seeds and soil, down to −300 MPa.
Answer: D
Explanation: The chamber needs a stalked leaf and stops near −8 MPa. It cannot measure single cells — that is the pressure probe.
18. Interactive quiz — test yourself
This quiz is separate from the MCQs above. All 20 questions are different. Select an answer for each question. You will not see whether an answer is correct until you submit all 20 questions. Use Reset to try again.
How did you do? If you scored 16 or more, you can explain psychrometry, the Kelvin equation and the thermocouple psychrometer confidently in an exam.
If you scored less than 16, re-read Sections 2, 3 and 5 before moving on to the next Plant Physiology post.
Continue learning — courses and related posts
- Water Potential: Components, Gradients and Measurement
- Water in the Soil and its Potentials
- Water Absorption in Plants
- Molecular Structure and Properties of Water
- The Soil-Plant-Atmosphere Continuum
- Water in Cells: Organelles, Biomolecules & Cellular Functions
- Abscisic Acid: A Comprehensive Exploration in Plant Physiology
- Salicylic Acid in Plants: Roles and Functions
- Aquaporins: Structure, Types, Regulation and Roles
- Turgor Pressure and the Höfler Diagram: How Plant Cells Stay Firm
- The Pressure Chamber (Scholander Pressure Bomb): How to Measure Leaf Water Potential
Next in this course: Pressure Probe: How to Measure Turgor Directly Inside One Living Cell
Sources
The textbook sources are the primary references for the definitions, methods and values in this post.
- Spanner DC. The Peltier effect and its use in the measurement of suction pressure. Journal of Experimental Botany. 1951;2:145–168.
- Richards LA, Ogata G. Thermocouple for vapor pressure measurement in biological systems. Science. 1958;128:1089–1090.
- Andraski BJ, Scanlon BR. Thermocouple psychrometry. In: Dane JH, Topp GC, editors. Methods of Soil Analysis, Part 4. Madison: Soil Science Society of America; 2002.
- Boyer JS. Measuring the Water Status of Plants and Soils. San Diego: Academic Press; 1995.
- Boyer JS. Leaf water potentials measured with a pressure chamber. Plant Physiology. 1967;42:133–137.
- Lang ARG. Osmotic coefficients and water potentials of sodium chloride solutions from 0 to 40 °C. Australian Journal of Chemistry. 1967;20:2017–2023.
- Brown RW, Bartos DL. A calibration model for screen-caged Peltier thermocouple psychrometers. USDA Forest Service Research Paper INT-293. 1982.
- Campbell GS. Measurement of water potential using thermocouple hygrometers. 1979. (Cited in Andraski & Scanlon 2002 for the temperature-error analysis.)
- Wescor, Inc. HR-33T Dew Point Microvoltmeter Instruction / Service Manual. 1978, 1986, 2001.
- METER Group. WP4C Dew Point PotentiaMeter — https://metergroup.com/products/wp4c/
- Miller ML, Estiarte E, Johnson DM, Zinnert J. Psychrometry for water potential measurements. ClimEx Handbook, section 5.8 — https://climexhandbook.w.uib.no/2019/11/03/psychrometry-for-water-potential-measurements/
- Taiz L, Zeiger E, Møller IM, Murphy A. Plant Physiology and Development, 6th edition. Sunderland: Sinauer Associates; 2015.
- Kirkham MB. Principles of Soil and Plant Water Relations, 2nd edition. San Diego: Academic Press; 2014.
Figures used in this post — credits and reuse
- Hero — A leaf disc sealed in a steel psychrometer chamber on a laboratory bench (photo, self-generated).
- Figure 1 — A thermocouple psychrometer: sample chamber, sensing junction, reference junction and microvoltmeter, labelled (diagram, self-generated).
- Figure 2 — A leaf disc in a sample chamber, the sample behind the Kelvin table (photo, self-generated).
- Figure 3 — A psychrometer calibration curve: microvolt output against known water potential, straight line, labelled (diagram, self-generated).
- Figure 4 — The dew-point junction: a cooled junction with a drop of dew, chamber air and sample below, labelled (diagram, self-generated).
- Figure 5 — Temperature error: how a small sample-sensor temperature gap shifts the apparent water potential, labelled (diagram, self-generated).
- Figure 6 — A leaf disc sealed in a sample chamber, ready for its reading (photo, self-generated).
- Figure 7 — A pressure chamber and a psychrometer side by side on a laboratory bench (photo, self-generated).








Courses
Diversity of Plants (BOT-303)
Pteridophytes & Gymnosperms (BOT-509)
Phycology & Bryology
Plant Tissue Culture
Plant Biochemistry II
Plant Physiology I
Plant Physiology II
Mycology & Plant Pathology
Microbiology & Virology
Cell Biology, Genetics & Evolution
Mycology (BOT-405)
Genetics I
Biochemistry
Plant Taxonomy & Systematics
Diversity of Vascular Plants
Advanced Plant Ecology
Plant Ecology II
Biodiversity & Conservation
Functional Biology
Botany Laboratory Techniques
Laboratory Techniques
Topics
Microbiology
Cell Biology
Practice
Study Tools
0 Comments
Have a biology question or spotted an error?
Send feedback to PreachBio →