Laboratory Glassware: Types, Accuracy and Measuring Volumes
Every measurement in biology starts with a vessel. If the vessel is wrong, the number is wrong, and no amount of careful maths afterwards can rescue it. This lesson teaches you the measuring family of laboratory glassware (vessels and tubes made of glass for laboratory work): what each piece is for, how exact it really is, how to read it, how to check it, and how to choose the right one — from a 2-litre beaker down to a 2-microlitre micropipette tip.
The habits in this lesson sit alongside the general rules in the lab best practices lesson and the handling rules in the glassware safety lesson. Those lessons cover working safely; this one covers working exactly — tolerances (the small allowed error printed on a vessel), percentage error, calibration (checking what a vessel really delivers against a trusted standard) and the standard solution (a solution whose concentration is known exactly), step by step.
Revision in one page — the numbers that matter
| Vessel or rule | Number | Why it matters |
|---|---|---|
| 100 mL volumetric flask, Class A tolerance | ±0.08 mL | Under 0.1% error at its full volume — the flask a standard solution is made in |
| 250 mL volumetric flask, Class A tolerance | ±0.12 mL | About 0.05% error — bigger flasks are proportionally even better |
| 25 mL volumetric (bulb) pipette, Class A tolerance | ±0.03 mL | Percentage error 0.12% — the most trusted fixed volume on the bench |
| 50 mL burette, Class A tolerance | ±0.05 mL | Each reading is read to ±0.05 mL, so a titre (two readings) carries ±0.10 mL |
| 100 mL measuring cylinder, Class A tolerance | ±0.5 mL | 0.5% at 100 mL — but 2.5% if you measure only 20 mL in it |
| Beaker scale | About ±5% | The numbers on a beaker are a rough guide for mixing, never a measurement |
| Concordant titres | Within 0.10 mL of each other | Only concordant titres are averaged; the rough titre is thrown away |
| Micropipette ranges | P20: 2–20 µL · P200: 20–200 µL · P1000: 200–1000 µL | Always pick the smallest pipette whose range covers your volume |
| Water for the weighing check | 1.00 g ≈ 1.00 mL (0.998 g/mL at 20 °C) | Weighing delivered water is how a vessel's true volume is checked |
Choosing the right vessel — the measuring family
Glassware falls into families by job. Some vessels contain a volume, some deliver a volume, and some are not measuring tools at all — they hold, mix or heat. The identification lesson on laboratory glassware names each piece and explains the "to contain" and "to deliver" marks; this section is about choosing between them, and the fuller bench survey in the laboratory apparatus lesson shows where they all live.
Beaker. A straight-sided open vessel for holding, mixing and heating. Its printed scale is a guide only (about ±5%), so nothing that matters is ever measured in a beaker. Its tall, narrow brother, the tall-form (Berzelius) beaker, is a heating and holding vessel too — it is not a titration vessel, whatever its shape suggests.
Conical (Erlenmeyer) flask. The titration vessel. WHAT: a flask with a wide flat base and a narrow neck. WHY: the wide base holds the liquid in a shallow pool while the narrow neck stops it splashing out as you swirl — and swirling without losing a drop is the whole skill of titration. It is also a safe boiling and storage shape, but its volume markings are rough, so it is not a measuring vessel either.
Measuring cylinder. A tall narrow tube on a base, graduated (marked with a scale of lines) all the way up. It measures everyday volumes — making up a water bath, rough dilutions — at middling accuracy. Its shape is its secret: the narrow bore stretches each millilitre over several millimetres of height, so the scale can be read finely.
Volumetric flask. A flat-bottomed flask with one single mark on a long narrow neck. It makes ONE exact volume of solution — 100.0 mL, 250.0 mL — and nothing else. It is the home of the standard solution, and the mark sits on the narrow neck because there a tiny extra volume moves the level a long way, so the mark can be hit exactly.
Volumetric (bulb) pipette. A tube with a bulb in the middle and one single mark. It delivers ONE exact volume — 10.00 mL, 25.00 mL — more accurately than anything else on the bench. Most of its volume sits in the bulb, so small level differences at the mark change the delivered volume by almost nothing.
Graduated pipette. A straight graduated tube that delivers variable (changing) volumes between its marks — the tool for in-between amounts like 2.5 mL, which no bulb pipette is made in. Less exact than a bulb pipette, far better than a cylinder at small volumes.
Burette. A long graduated tube with a tap at the bottom, clamped vertically. It delivers a measured, changing volume — the titre (the volume of solution delivered in a titration) — from the difference between two readings. Its scale runs 0 at the top to 50 at the bottom because the level falls as it delivers.
Test tube and boiling tube. Small-scale reaction and observation vessels. A boiling tube is the heating member of the pair: shorter, wider and thicker-walled than a test tube, so a small volume can be heated directly in it without the liquid surging out.
Accuracy, precision and tolerance — how exact is exact?
Two words that sound alike mean different things. Accuracy is closeness to the true value. Precision is agreement between repeated measurements. A pipette that delivers 24.80, 24.81, 24.79 and 24.80 mL when it should deliver 25.00 mL is precise — its readings agree beautifully — but it is NOT accurate, because every reading sits 0.20 mL below the truth. Good practical work needs both, and the fix for each is different: precision improves with careful technique, accuracy improves only by checking the instrument — calibration, below.
Every vessel carries a tolerance: the largest error the maker guarantees it will stay inside, printed on the glass with its class. Class A is the accurate grade; Class B allows roughly twice the error. A tolerance is a fixed number of millilitres, so what it does to your result depends on the volume you measure. The tool for judging that is the percentage error:
Percentage error = tolerance ÷ volume measured × 100
- Worked example 1. The 25 mL bulb pipette, tolerance ±0.03 mL: 0.03 ÷ 25 × 100 = 0.12%. Tiny — this is why titrations trust it.
- Worked example 2. The 100 mL cylinder, tolerance ±0.5 mL, used at its full 100 mL: 0.5 ÷ 100 × 100 = 0.5%. Acceptable for rough work.
- Worked example 3. The same cylinder used to measure 20 mL: the tolerance is still ±0.5 mL, so 0.5 ÷ 20 × 100 = 2.5%. The vessel did not change — the share did.
Worked example 3 teaches the smallest-vessel rule: always use the smallest vessel that holds the volume you need, so the fixed tolerance is the smallest possible share of it. The mirror-image rule also holds: never use a vessel so small that the job needs several fillings — each filling adds its own error and its own spillage risk.
One more marking matters. Vessels are calibrated either TC ("to contain" — the marked volume is what is inside, like a volumetric flask) or TD ("to deliver" — the marked volume is what runs out, like a pipette or burette, with the film that stays behind already allowed for). The glassware safety lesson works through TC and TD, the two classes and drainage times in detail; here, just remember that you read a TC vessel by what it holds and a TD vessel by what left it.
Reading the meniscus — eye level, the right edge
The surface of a liquid in a narrow tube is not flat. It curves up at the edges into a shallow bowl called the meniscus (the curved surface of a liquid in a tube). Reading a volume means reading that curve correctly, and three rules cover almost every case:
- Eye level. Bring your eye exactly level with the meniscus before you read. Reading from above or below gives parallax error (a false reading caused by viewing a scale from an angle) — the identification lesson on laboratory glassware shows the geometry of it.
- Read the bottom of the meniscus for ordinary colourless solutions — the bottom of the bowl is the level the scale was calibrated against.
- Read the top for dark solutions. Potassium manganate(VII) is so deep a purple that the bottom of its meniscus cannot be seen at all, so the top edge is read instead — the same edge for the start and finish readings, so a difference (a titre) stays correct.
Two helps are worth the habit. Hold a white card (a dark card for pale liquids) behind the tube: against a plain, even background the meniscus shows as a sharp dark edge instead of getting lost against the bench. And give the liquid a moment to settle and drain down the walls before reading — a reading taken while the walls are still wet above the surface is a reading of the wrong level.
Burette work — delivered volumes and honest titres
A burette never has to start at zero. It measures what it delivers, and a delivered volume is a difference: titre = final reading − initial reading. If the meniscus starts at 12.35 mL and ends at 35.60 mL, the volume delivered is 35.60 − 12.35 = 23.25 mL. Running solution out just to reach exactly 0.00 wastes reagent and adds handling errors; record the start wherever it sits, as long as it sits below the 0 mark's zero and you can read it clearly.
Before the first reading, one small job matters: take the filling funnel out. Liquid clings to a funnel and its stem after filling, and drops that slip down later change the volume inside the burette after you have read it — the readings then stop describing what was delivered. Fill, remove the funnel, wait a moment, then read. The burette is also conditioned (rinsed through with the solution it is about to hold) before filling, so the titrant is not diluted by rinse water left in the tube — the pre-use checks, including the air bubble in the tip and the vertical clamp, are worked through in the laboratory glassware lesson.
Titrations are repeated for a reason. The first titre is a rough titre: a scout that finds roughly where the end point (the moment the reaction is just complete, shown by the indicator) lies. Later titres, done slowly near that point, should agree — titres within 0.10 mL of each other are called concordant, and only concordant titres are averaged. Worked example: rough 24.10 mL (thrown away), then 23.15, 23.25 and 23.20 mL — all within 0.10 mL, so the mean titre is (23.15 + 23.25 + 23.20) ÷ 3 = 23.20 mL.
Finally, be honest about the uncertainty. Each burette reading is good to about ±0.05 mL, and a titre uses two readings, so the titre's uncertainty is ±0.05 + ±0.05 = ±0.10 mL. On a 20.00 mL titre that is 0.10 ÷ 20.00 × 100 = 0.5% — small, but real, and it is why concentrations from titrations are quoted to three or four significant figures, not ten.
Making a standard solution — step by step
A standard solution is the solution every other result in an analysis is compared against, so its concentration must be exactly the one you calculated. It is made in a volumetric flask, and the method below is the one the examination boards' practical endorsements expect. Notice what the method never does: the solid is never dissolved inside the flask itself. The flask's narrow neck makes stirring hard, solid can lodge in the neck out of sight, and once the flask stands at its mark there is no room left to fix anything.
- Weigh the solid. WHAT: the exact mass the calculation asks for. HOW: weigh a clean, dry beaker (or weighing boat) on an analytical balance, add the solid, and record the mass to the balance's full precision. WHY: the concentration of the finished solution is built on this one number — an error here is in every drop.
- Dissolve it in the beaker. WHAT: get every crystal into solution before anything moves. HOW: add distilled water to roughly half the final volume, and stir with a glass rod until no crystal remains. WHY: undissolved solid transferred to the flask may never dissolve once the flask is full, and the concentration would be wrong by whatever is left behind. Distilled water, not tap water: tap water carries dissolved ions (charged particles) that can react with the solute or add unknown amounts of the very substances being studied.
- Transfer every drop. WHAT: move the solution — and everything dissolved in it — into the flask. HOW: pour through a funnel into the flask neck, then rinse the beaker, the rod and the funnel with distilled water, adding every rinsing to the flask. WHY: a film of solution clings to each surface, and that film holds some of the weighed solute. Rinsing it in makes the transfer quantitative (complete — everything weighed ends up in the flask), so the concentration you calculated is the concentration you have.
- Fill to the mark. WHAT: bring the volume to exactly the flask's one volume. HOW: add distilled water until the level nears the mark, then add it drop by drop with eye level at the mark, until the bottom of the meniscus sits exactly on the line. WHY: the flask is calibrated at that line and no other. And if the level creeps past the mark, there is no rescue — removing liquid removes an unknown share of solute from an unmixed solution. The solution is discarded and made again.
- Stopper and mix. WHAT: make the concentration the same everywhere in the flask. HOW: stopper the flask and invert it several times. WHY: fresh water layered on top of denser solution mixes only slowly on its own; an unmixed flask holds weak solution at the top and strong solution at the bottom, and the first sample poured out would not be the standard at all.
Watch the steps — making a standard solution
Step through the method. Each step names what happens, and Reset returns you to the start.
Make one standard solution the accurate way — weigh the solid, dissolve it, transfer every drop, fill to the mark, then stopper and mix.
Small volumes — micropipettes and friends
Below a millilitre, glass tubes give way to instruments built for microlitres (µL — millionths of a litre). The same choosing rule applies, in miniature.
Micropipette. An adjustable piston pipette with a disposable tip. The three standard sizes are the P20 (2–20 µL), the P200 (20–200 µL) and the P1000 (200–1000 µL). WHAT matters: choose the smallest pipette whose range covers your volume, because each one is least accurate at the bottom of its range. For 150 µL, the P200 (where 150 µL sits at 75% of its span) beats the P1000 (where it sits at 15%). HOW it works, step by step:
- Set the volume and fit a clean tip firmly.
- Press the plunger to the first stop only, then put the tip just 2–3 mm under the liquid surface. Deeper dipping coats the outside of the tip, and that coating is carried across as extra, unmeasured volume.
- Release the plunger slowly. A snapping release traps bubbles and can splash liquid up into the barrel — both leave the volume short.
- To deliver, press to the first stop, then on to the second stop, which blows out the last film clinging inside the tip.
- For thick, viscous (slow-flowing) liquids, use reverse pipetting: draw up to the second stop (more than the set volume), then dispense only to the first stop. The film that stays on the tip wall comes out of the extra you drew up, not out of the measurement.
Pasteur (dropping) pipette. An ungraduated glass or plastic tube with a rubber bulb. It moves drops — indicator, a top-up, a transfer — and measures nothing at all. Any volume that matters must pass through graduated or volumetric ware.
Bottle-top dispenser. Screwed onto a stock bottle, it delivers the same set volume stroke after stroke — fast, and kind to corrosive reagents that would otherwise be pipetted by hand. Two disciplines: prime it before first use (its feed tube starts full of air, and air-filled strokes deliver part-air shots that come out short), and check its calibration like any other measuring device.
Density bottle and hydrometer. Two ways to find a liquid's density (mass per unit volume). The density bottle (specific-gravity bottle) has a stopper pierced with a fine hole, so it always holds exactly the same volume: weigh it filled, subtract the empty mass, divide by the fixed volume — an accurate density. The hydrometer floats in the liquid and is read off its stem: it floats higher in a denser liquid, because it needs to displace less of it to support its own weight. Quick, but less accurate than the bottle.
Cuvette. The small cell that holds a sample in a colorimeter (an instrument that measures how much light a solution absorbs). It is measuring glassware too: light passes through its two clear faces, so fingerprints there absorb and scatter light and push the absorbance reading up. Wipe the clear faces, hold it by the frosted sides, and put it in the same way round every time — no two faces are perfectly identical, and one fixed orientation makes any small unevenness affect every reading equally. Sample preparation of this care sits next to the measurement habits in the microscopy and cell size lesson.
Calibration — proving a vessel tells the truth
Calibration is a check, not a wash: the volume a vessel really delivers or contains is measured against a trusted standard and compared with the value printed on it. The standard school and university method is gravimetric (based on weighing):
- Fill the vessel with distilled water exactly as you would in real use.
- Deliver (or pour out) the water into a tared (zeroed with its container) vessel on a balance, and record the mass.
- Convert mass to volume: at 20 °C water's density is 0.998 g/mL, and the school rule 1.00 g ≈ 1.00 mL is close enough for a pass/fail judgement.
- Compare with the printed volume. The difference must lie inside the vessel's tolerance.
Worked example: a 25 mL bulb pipette delivers water massing 24.86 g — about 24.86 mL, which is 0.14 mL short of 25.00 mL. Its Class A tolerance is ±0.03 mL, so it fails by nearly five times the allowance and must not be used for accurate work. A pipette that passes, on the other hand, has earned its printed value. Balances earn theirs the same way — the record-keeping side of checking is covered in the lab best practices lesson, and the mass standards themselves are the subject of the triple-beam balance lesson. Clean glassware is part of honest measurement too: grease makes water bead instead of draining in a smooth film, and a beading vessel delivers the wrong volume, which is why the glassware cleaning discipline and the storage rules in treatment of glassware after washing belong in this lesson's family.
Why glass? — and the vessels built for heat
Plastic ware is cheap and survives being dropped, yet volumetric ware is still glass. The reasons are physical: glass is rigid (a plastic wall flexes in the hand and the volume changes with it), it takes fine etched marks, it resists almost every laboratory chemical, and borosilicate glass (the heat-resistant laboratory glass) expands very little with temperature. Plastic keeps its proper jobs — wash bottles, rough cylinders, storage — but its tolerances are wider, and wide tolerances are exactly what volumetric work cannot use.
Heat is the other place material matters. Borosilicate's low expansion is what lets a boiling tube take a direct flame where ordinary glass would crack from thermal shock (cracking caused by uneven heating); the physics of that, and the inspection rules for chips and cracks, are covered in the laboratory techniques lesson. Treat every vessel in this lesson as unfit for measuring the moment it is chipped — a chipped rim or neck changes the volume and sheds fragments, and broken glass goes straight to the sharps stream described in the lab waste management lesson.
Never do this
- Never measure a volume in a beaker or a conical flask — their scales are guides, not measurements.
- Never dissolve a solid for a standard solution inside the volumetric flask.
- Never rescue an overshot flask by pipetting liquid back out — discard it and remake it.
- Never leave the filling funnel in a burette while you read or titrate.
- Never average a rough titre in with concordant titres.
- Never press a micropipette plunger to the second stop while drawing liquid up.
- Never use a vessel — or a micropipette — outside the range it was made for.
- Never trust a printed volume on a vessel that has failed its calibration check.
Common mistakes — and the fix
| Mistake | Why it is wrong | The fix |
|---|---|---|
| "Tall-form beakers are for titrations" | A beaker cannot be swirled without spilling, and it measures nothing; a tall form changes neither fact | Titrate in a conical flask; tall-form beakers are for heating and holding |
| Measuring 20 mL in a 100 mL cylinder | The fixed ±0.5 mL tolerance becomes a 2.5% error at 20 mL | Smallest-vessel rule — a 25 mL cylinder, or a pipette if the volume must be exact |
| Reading the bottom of a dark meniscus by guessing | The bottom edge is invisible in a dark solution, so the guess is pure error | Read the top of the meniscus — the same edge for both readings |
| Chasing 0.00 mL at the start of a titration | A titre is a difference; the starting value cancels out | Record the start reading wherever the meniscus sits and subtract |
| Skipping the washings when transferring a solution | The film left in the beaker holds weighed solute, so the flask ends up weaker than calculated | Rinse beaker, rod and funnel into the flask — quantitative transfer |
| Pouring from a flask that was never inverted | Unmixed solution is weak at the top and strong at the bottom | Stopper and invert several times before taking any sample |
| Using a P1000 for 150 µL | Micropipettes are least accurate at the bottom of their range — 150 µL is 15% of a P1000's span | Use the P200, where 150 µL sits at 75% of the span |
Questions students ask
Can I measure with a conical flask if I am careful? No. Care does not narrow a tolerance: the flask's wide body spreads its rough scale over a tiny height change, so even a perfect reading carries a large error. Flask for swirling, cylinder for rough volumes, pipette or burette for exact ones.
Why does the burette not need to start at zero? Because you never use either reading on its own — only their difference, and the starting value cancels in the subtraction. Any readable starting level gives the same titre.
My repeats agree with each other but not with the expected value. What is broken? Your precision is fine and your accuracy is not — the classic signature of an instrument fault: an uncalibrated pipette, a wrongly read meniscus edge, or a vessel used outside its range. Check the instrument (gravimetric check) before blaming your technique.
Is a measuring cylinder with a single mark a thing? No — cylinders are graduated all the way up; that is their job. The single-mark vessels are the volumetric flask (contains one volume) and the bulb pipette (delivers one volume). If a question describes a "volumetric cylinder" with one mark, it has mixed the two up.
How often should glassware be calibration-checked? Volumetric ware in regular accurate use is checked on a schedule (termly in a school lab, and always after any repair, chip or suspicious result), and the result is recorded. The glassware cleaning MCQs lesson covers the care routine that keeps a checked vessel honest between checks.
20 practice MCQs — answers explained
20 Explained MCQs — answers hidden until opened
Open each question to reveal its answer and explanation. The interactive quiz below uses a different set of questions.
1. A pipette is used four times and delivers 24.80 mL, 24.81 mL, 24.79 mL and 24.80 mL, when it should deliver 25.00 mL. How should these readings be described?
A. Accurate but not precise — the readings average out at the true value B. Neither accurate nor precise — the readings are all different C. Both accurate and precise — the spread is very small D. Precise but not accurate — the readings agree with each other but all sit below the true value
Answer: D. Precision is about agreement between repeats: these four readings cluster within 0.02 mL, so they are precise. Accuracy is about closeness to the true value: they all sit about 0.20 mL below 25.00 mL, so the pipette is not accurate — it needs its calibration checked.
2. A 25 mL Class A volumetric pipette has a tolerance of ±0.03 mL. What is its percentage error when it delivers its full 25 mL?
A. 0.12% B. 3% C. 1.2% D. 0.03%
Answer: A. Percentage error = tolerance ÷ volume × 100 = 0.03 ÷ 25 × 100 = 0.12%. This tiny percentage is why a volumetric pipette is trusted for the fixed volume in a titration.
3. A 100 mL Class A measuring cylinder has a tolerance of ±0.5 mL. A student uses it to measure out 20 mL. What is the percentage error, and what rule does it teach?
A. 2.5% — large cylinders must be read twice and averaged B. 2.5% — use the smallest vessel that holds the volume you need C. 0.1% — the tolerance shrinks for smaller volumes D. 0.5% — a Class A vessel always gives 0.5% error
Answer: B. The tolerance stays ±0.5 mL whatever volume is in the cylinder, so at 20 mL the percentage error is 0.5 ÷ 20 × 100 = 2.5%. The same cylinder at its full 100 mL gives only 0.5%. Rule: choose the smallest vessel that holds the volume, so the fixed tolerance is the smallest possible share of it.
4. A 25 mL volumetric pipette is checked by weighing the water it delivers (take 1.00 g of water as 1.00 mL). The water masses 24.86 g. The pipette's tolerance is ±0.03 mL. What does the check show?
A. It delivers 25.14 mL, slightly too much B. It delivers about 24.86 mL — 0.14 mL short, far outside its tolerance, so it fails the check C. Nothing — a balance cannot check a volume D. It passes — 24.86 g is within ±0.03 mL of 25 mL
Answer: B. Weighing delivered water is the standard gravimetric check. 24.86 g of water is about 24.86 mL, which is 0.14 mL below the marked 25.00 mL — nearly five times the ±0.03 mL tolerance. The pipette must not be used for accurate work until it is repaired or replaced.
5. Potassium manganate(VII) solution is so dark that the bottom of its meniscus cannot be seen in a burette. How is its level read?
A. Read the top of the meniscus instead B. Read the middle of the dark band C. Shake the burette until the colour lightens D. Estimate the level from the tap upwards
Answer: A. For colourless solutions you read the bottom of the concave meniscus. A dark solution hides that bottom edge completely, so the standard rule is to read the top of the meniscus — the same edge for both readings, so the difference (the titre) stays correct.
6. A burette is filled through a funnel. Why must the funnel be taken out before the first reading is recorded?
A. The funnel blocks the view of the tap B. Filling funnels react with titrants C. Drops clinging to the funnel can fall in later and change the volume after the reading D. The funnel makes the burette too heavy for the clamp
Answer: C. Liquid clings to the funnel and its stem after filling. If the funnel stays in place, those drops can slip down into the burette during the titration, so the volume delivered no longer matches the difference between the two readings. Fill, remove the funnel, then read.
7. At the start of a titration the burette meniscus sits at 7.35 mL, not at 0.00 mL. What should the student do?
A. Start again with a different burette B. Record 7.35 mL as the initial reading and carry on — a titre is a difference between two readings C. Add 7.35 mL to the final reading at the end D. Run solution out until the level is exactly 0.00 mL before starting
Answer: B. A burette measures the volume DELIVERED, which is final reading minus initial reading. Any clear starting level works, because the starting value cancels out in the subtraction. Chasing exactly 0.00 wastes solution and adds errors of its own.
8. Titres of 23.15 mL, 23.25 mL and 23.20 mL are obtained after a rough titre of 24.10 mL. Why are the three later titres called concordant?
A. They are all larger than the rough titre B. Their average is a whole number of millilitres C. They agree with each other to within 0.10 mL D. They were all read by the same person
Answer: C. Concordant means reproducible: titres within 0.10 mL of each other show the technique is under control. Only concordant titres are averaged (here the mean is 23.20 mL). The rough first titre is a scout — it only tells you roughly where the end point lies and is never included.
9. Each burette reading is uncertain by about ±0.05 mL. A titre is worked out from two readings. What is the uncertainty in the titre?
A. ±0.10 mL — the uncertainties of the two readings add B. ±0.025 mL — the uncertainties partly cancel C. ±0.05 mL — only the final reading matters D. Zero — subtraction removes uncertainty
Answer: A. The starting reading could be 0.05 mL out one way and the final reading 0.05 mL out the other way, so the difference can be wrong by up to 0.05 + 0.05 = 0.10 mL. On a 20.00 mL titre that is a 0.5% uncertainty — worth knowing before a concentration is claimed to be exact.
10. A solid is dissolved in a beaker, then the solution is poured through a funnel into a volumetric flask. Why are the beaker, the stirring rod and the funnel then rinsed with distilled water INTO the flask?
A. To wash the funnel before it is put away B. To cool the solution down before it is made up to the mark C. To make the flask easier to stopper D. So that every trace of the solute reaches the flask — the transfer must be quantitative
Answer: D. A film of solution clings to the beaker, rod and funnel, and that film contains some of the weighed solute. Rinsing it into the flask makes the transfer quantitative — all of the solid you weighed ends up in the flask, so the concentration you calculated is the concentration you actually have.
11. A volumetric flask has been filled exactly to its mark. Why must it be stoppered and inverted several times before any solution is taken out of it?
A. Inverting checks that the stopper does not leak B. Air bubbles raise the level above the mark C. Without mixing, the solution is stronger at the bottom than at the top D. The mark is only correct while the flask is upside down
Answer: C. Fresh water added at the top mixes only slowly with the denser solution below it, so an unmixed flask holds weak solution at the top and strong solution at the bottom. Stoppering and inverting several times makes the concentration the same throughout — only then is any sample taken from it the standard solution.
12. When a micropipette is used to deliver its set volume, the plunger is pressed to the second stop right at the end of dispensing. What does the second stop do?
A. It draws extra liquid into the tip B. It blows out the last of the liquid clinging inside the tip C. It doubles the volume delivered D. It locks the volume dial
Answer: B. The first stop dispenses the measured volume. A thin film still clings inside the tip; pressing on to the second stop pushes a puff of air through that blows the rest out, so the full set volume is delivered. When DRAWING liquid up, though, you press only to the first stop.
13. You need 150 µL of an enzyme solution. A P200 (20–200 µL) and a P1000 (200–1000 µL) micropipette are both on the bench. Which should you use, and why?
A. Either — both can physically hold 150 µL B. The P200 — a micropipette is most accurate in the upper part of its range, and 150 µL sits high in the P200's range C. Neither — 150 µL must be measured in a measuring cylinder D. The P1000 — a bigger pipette is always more accurate
Answer: B. Every micropipette is least accurate at the bottom of its range. 150 µL is 75% of the P200's span but only 15% of the P1000's span, so the P200 delivers it far more reliably. Rule: choose the smallest micropipette whose range covers the volume.
14. What is a Pasteur (dropping) pipette able to do?
A. Measure volumes to the nearest 0.1 mL B. Replace a micropipette for microlitre work C. Move drops and small rough amounts of liquid from place to place — it cannot measure a volume D. Deliver exactly 1.00 mL each time it is squeezed
Answer: C. A Pasteur pipette has no graduations at all. It is a transfer tool — adding indicator drop by drop, topping up, moving a little liquid — and nothing it delivers is a known volume. Any volume that matters must be measured with graduated or volumetric ware.
15. How does a density bottle (specific-gravity bottle) give the density of a liquid?
A. It changes colour according to the density B. It measures how fast the liquid flows out of the bottle C. It floats in the liquid and is read like a thermometer D. Its stopper has a fine hole, so the bottle always holds exactly the same volume; density = the mass of liquid filling it ÷ that fixed volume
Answer: D. The pierced stopper lets excess liquid escape, so the bottle is filled to exactly its marked volume every time. Weigh the filled bottle, subtract the empty mass, and divide by the fixed volume: that quotient is the density, found far more accurately than with a floating hydrometer.
16. Why is a measuring cylinder made tall and narrow instead of short and wide?
A. In a narrow tube each millilitre spans a tall slice of height, so the scale is stretched out and can be read finely B. The narrow shape stops the liquid evaporating C. A tall cylinder cannot be knocked over D. Narrow cylinders are cheaper to blow from glass
Answer: A. Volume divided by cross-section gives height. A narrow bore turns 1 mL into several millimetres of height change, spreading the graduations apart so readings are finer. The same trick appears in the narrow neck of a volumetric flask and the thin tube of a burette.
17. Plastic measuring ware is cheap and unbreakable. Why is volumetric glassware still made of glass?
A. Plastic dissolves in water B. Glass is rigid, takes fine etched marks, resists laboratory chemicals and expands very little — plastic walls flex and its tolerances are wider C. Glass is cheaper than plastic D. Plastic cannot be made transparent
Answer: B. A volumetric vessel must hold its calibrated volume exactly. Glass does not flex when squeezed or warmed by a hand, it can be etched with very fine marks, it stands up to almost every reagent, and borosilicate glass barely expands with temperature. Plastic is fine for wash bottles and rough cylinders, but its flexing walls and wider tolerances rule it out for volumetric work.
18. Before a cuvette goes into a colorimeter, its two clear faces are wiped and it is afterwards held only by the frosted sides. Why?
A. Wiping charges the glass with static B. The frosted sides are where the scale is printed C. Fingerprints and grease on the clear faces scatter and absorb light, which falsifies the absorbance reading D. The clear faces are thinner and might crack
Answer: C. The light beam passes through the clear faces. A fingerprint film absorbs and scatters some of that light, and the colorimeter cannot tell film from sample — the absorbance reads too high. Handle cuvettes by the frosted sides, wipe the optical faces, and put the cuvette in the same way round every time.
19. How does a boiling tube differ from an ordinary test tube?
A. It has a ground-glass stopper fitted B. It is graduated in millilitres C. It is made of plastic instead of glass D. It is shorter, wider and thicker-walled, so a small volume can be heated directly in it
Answer: D. A boiling tube is built for heat: its wide bore leaves room for boiling liquid to surge without ejecting, and its thicker wall stands up to direct heating better. An ordinary test tube is a small-scale reaction and observation vessel, not a heating vessel for liquids.
20. What is the job of a bottle-top dispenser?
A. It filters the reagent as it is poured B. It stores volatile reagents without evaporation C. It screws onto a reagent bottle and delivers the same set volume again and again, without refilling a pipette each time D. It measures the density of the reagent in the bottle
Answer: C. Each stroke of a bottle-top dispenser delivers the volume set on it, straight from the stock bottle. That makes repeated dispensing fast and keeps corrosive reagents off pipettes and hands. It still has to be primed before first use and its calibration checked, like any measuring device.
Interactive quiz — 20 questions
Separate Interactive Quiz — 20 different questions
These quiz questions are different from the MCQs above. Select an answer for each question. You will not see whether an answer is correct until you submit all 20 questions. Use Reset to try again.
Sources consulted
AQA GCSE Biology (8461) and A-level Biology (7402) specifications and required practical handbook; OCR GCSE Biology A (J247) and A-level Biology A (H432) Practical Activity Group briefs; Edexcel GCSE Biology (1BI0) and A-level Biology B (9BN0) specifications; College Board AP Biology Course and Exam Description (Science Practices); IB Biology Guide (first assessment 2025); OpenStax Biology 2e; St Mary's College laboratory glassware appendix (Class A tolerances); IGNOU titrimetry unit (burette and pipette practice); Biology LibreTexts quantitative techniques appendix (micropipette stops and immersion depth). Save My Exams, BBC Bitesize and Khan Academy were compared for coverage before writing: their glassware content sits inside chemistry practical units (titration technique) or theory units (microscopy calculations), and none of them teaches volumetric accuracy — tolerances, percentage error, calibration — as a biology lesson, which is the gap this lesson fills. Syllabus requirements checked November 2026.
Continue learning
- Laboratory glassware — naming every vessel, TC and TD marks, and the parallax geometry.
- Glassware safety — Class A and Class B, drainage times, and safe handling, in 20 MCQs.
- Glassware cleaning techniques — washing, conditioning and drying, so clean glass measures true.
- Radiation hazards — the other exact-measurement discipline of the biology laboratory.
- Microscopy and cell size — measuring the very small, with a calibrated scale.